There are two types of coulometric reagents: the anolyte (generating solution),which is placed in the anode chamber of the electrolysis cell, and the catholyte (counter electrolyte), which is placed in the cathode chamber. There are also special anolyte for use with ketones, lower carboxylic acids and silicone oils. AQUAMICRON™ can be used in coulometric moisture measurement systems sold by various companies, and it has a worldwide reputation for excellent performance.
Example (Coulometric Reagents)AQUAMICRON™ Series
Product name | Use | Specification | Package | |
---|---|---|---|---|
AQUAMICRON™ |
AXI | *AXI [For frit or fritless type cell, equivalent to FLS] *AXI, AX & AX01[For General-use] Organic solvents, Inorganic chemicals, Oiles, Petroleum products, Various kind of gases, etc. |
Moisture maximum 0.15 mg/ml |
500mL |
AX | ||||
AX01 | 100mL | |||
AS | [For Oils] Naphtha, Gasolone, Diesel oil, Electrical insulation oil, etc. |
Moisture maximum 0.15 mg/ml |
500mL | |
AKX | [For Ketones] Ketones. Silicone oils. Low carboxylic acids, etc. |
Moisture maximum 0.15 mg/ml |
500mL | |
CXU | [Catholyte] Both AX and AS and AKX available to combine with |
Moisture maximum 0.6 mg/ml |
5mL x10/case |
|
FLS | [For fritless type cell] Organic solvents, Inorganic gases, etc. |
Moisture maximum 0.15 mg/ml |
500mL |
AQUAMICRON™ AXI/CXU or AX/CXU
Use | General samples | |
---|---|---|
Characteristics | Minimal environmental contamination | Does not include carbon tetrachloride or chloroform. |
Wide range of applications | Suitable for petroleum products Can be used with moisture vaporization method (When using AQUAMICRON™ AX or AXI in a moisture vaporization method, replenish the vaporized portion with methanol.) |
|
High quality, high performance | Excellent reputation for accurate moisture measurement and end-point stability Approximately 800mg of moisture can be measured per 100mL of AQUAMICRON™ AX or AXI, and approximately 150mg per 5mL of AQUAMICRON™ CXU |
|
Easy to use | These products can be used in the existing coulometric moisture measurement system. |
AQUAMICRON™ AS/CXU
Use | General samples | |
---|---|---|
Characteristics | Wide range of applications | Especially suitable for oils, petroleum products |
High quality, high performance | Excellent reputation for accurate moisture measurement and end-point stability. Approximately 800mg of moisture can be measured per 100mL of AQUAMICRON™ AS, and approximately 150mg per 5mL of AQUAMICRON™ CXU. |
|
Easy to use | These products can be used in the existing coulometric moisture measurement system. |
AQUAMICRON™ AKX/CXU
Use | Ketones, lower carboxylic acids, some aldehydes (aromatics) | |
---|---|---|
Characteristics | Minimal environmental contamination | Does not include chloroform or methyl cellosolve |
Wide range of applications | Suitable for ketones, lower carboxylic acids and some aldehydes | |
High quality, high performance | Accurate moisture measurement, good end-point stability Approximately 300mg of moisture can be measured per 100ml of AQUAMICRON™ AKX, and approximately 100mg per 5mL of AQUAMICRON™ CXU (when used with AQUAMICRON™ AKX). |
|
Easy to use | These products can be used in the existing coulometric moisture measurement system. |
AQUAMICRON™ FLS
Use | Suitable for alcohols, esters, benzene, toluene, inorganic gases, etc. | |
---|---|---|
Characteristics | Easy to use | AQUAMICRON™ AXI or FLS is a single-solution type reagent. It should be used in single-solution electrolytic cells without diaphragm, fritless type cell. Can be used with moisture vaporizati on method(When using AQUAMICRON™ AXI or FLS in a moisture vaporization method, replenish the vaporized portion with methanol.) |
Easy to use | Ceramic diaphragm is not used, therefore it is easy to maintain electrolytic cell in good condition. |